Raman probe for the simultaneous measurement of anion concentration in mixtures of salt solutions - Laboratoire Matériaux Optiques, Photonique et Systèmes Accéder directement au contenu
Communication Dans Un Congrès Année : 2013

Raman probe for the simultaneous measurement of anion concentration in mixtures of salt solutions

Résumé

We propose to detect by Raman spectrometry the presence of several anions in mixtures of salt solutions and to measure simultaneously their espective concentrations. In this study we show how chemometric methods based on statistical analysis like Principal Components Analysis (PCA) and Partial Least Square Regression (PLSR) can be applied for the spectra treatment. Unlike current water probes often based on conductivity, our new Raman sensor is able to identify simultaneously and in one single measurement several anions such as NO3- , SO42- , HPO42- or Cl - sought for the water pollution monitoring. Moreover our sensor gives rise to reliable, accurate and fast measurement (a few seconds) and can therefore be very useful for in situ water quality control applications.
Fichier principal
Vignette du fichier
2013 - Kauffmann.pdf (302.56 Ko) Télécharger le fichier
Origine : Fichiers produits par l'(les) auteur(s)
Loading...

Dates et versions

hal-01280731 , version 1 (12-05-2017)

Identifiants

  • HAL Id : hal-01280731 , version 1

Citer

T.H. Kauffmann, K. Ben Mabrouk, M.D. Fontana. Raman probe for the simultaneous measurement of anion concentration in mixtures of salt solutions. IEEE Sensors 2013, Nov 2013, Baltimore, United States. pp.671-674. ⟨hal-01280731⟩