Compressed Sensing Dynamic Reconstruction in Rotational Angiography

Abstract : This work tackles three-dimensional reconstruction of tomographic acquisitions in C-arm-based rotational angiography. The relatively slow rotation speed of C-arm systems involves motion artifacts that limit the use of three-dimensional imaging in interventional procedures. The main contribution of this paper is a reconstruction algorithm that deals with the temporal variations due to intra-arterial injections. Based on a compressed-sensing approach, we propose a multiple phase reconstruction with spatio-temporal constraints. The algorithm was evaluated by qualitative and quantitative assessment of image quality on both numerical phantom experiments and clinical data from vascular C-arm systems. In this latter case, motion artifacts reduction was obtained in spite of the cone-beam geometry, the short-scan acquisition, and the truncated and subsampled data.
Type de document :
Communication dans un congrès
MICCAI'12, Oct 2012, Nice, France. 7510, pp.223-230, 2012, 〈10.1007/978-3-642-33415-3_28〉
Liste complète des métadonnées

https://hal-supelec.archives-ouvertes.fr/hal-00756119
Contributeur : Alexandra Siebert <>
Soumis le : jeudi 22 novembre 2012 - 15:16:54
Dernière modification le : vendredi 8 février 2019 - 09:16:03

Lien texte intégral

Identifiants

Citation

Hélène Langet, Cyril Riddell, Yves Trousset, Arthur Tenenhaus, Elisabeth Lahalle, et al.. Compressed Sensing Dynamic Reconstruction in Rotational Angiography. MICCAI'12, Oct 2012, Nice, France. 7510, pp.223-230, 2012, 〈10.1007/978-3-642-33415-3_28〉. 〈hal-00756119〉

Partager

Métriques

Consultations de la notice

695